Evaluation of the Chlorophyll/Fluorescence Sensor of the YSI Multiprobe: Comparison to an Acetone Extraction Procedure

PDF Version Also Available for Download.

Description

The purpose of this study was to examine the suitability of the YSI model 6600 Environmental Monitoring System (multiprobe) for long term deployment at a site in Lewisville Lake, Texas. Specifically, agreement between a laboratory extraction procedure and the multiprobe chlorophyll/fluorescence readings was examined. Preliminary studies involved determining the best method for disrupting algal cells prior to analysis and examining the precision and linearity of the acetone extraction procedure. Cell disruption by mortar and pestle grinding was preferable to bath sonication. Comparison of the chlorophyll/fluorescence readings from the multiprobe and the extraction procedure indicated that they were significantly correlated but … continued below

Creation Information

Lambert, Patricia May 2001.

Context

This thesis is part of the collection entitled: UNT Theses and Dissertations and was provided by the UNT Libraries to the UNT Digital Library, a digital repository hosted by the UNT Libraries. It has been viewed 1455 times. More information about this thesis can be viewed below.

Who

People and organizations associated with either the creation of this thesis or its content.

Chair

Committee Members

Publisher

Rights Holder

For guidance see Citations, Rights, Re-Use.

  • Lambert, Patricia

Provided By

UNT Libraries

The UNT Libraries serve the university and community by providing access to physical and online collections, fostering information literacy, supporting academic research, and much, much more.

Contact Us

What

Descriptive information to help identify this thesis. Follow the links below to find similar items on the Digital Library.

Description

The purpose of this study was to examine the suitability of the YSI model 6600 Environmental Monitoring System (multiprobe) for long term deployment at a site in Lewisville Lake, Texas. Specifically, agreement between a laboratory extraction procedure and the multiprobe chlorophyll/fluorescence readings was examined. Preliminary studies involved determining the best method for disrupting algal cells prior to analysis and examining the precision and linearity of the acetone extraction procedure. Cell disruption by mortar and pestle grinding was preferable to bath sonication. Comparison of the chlorophyll/fluorescence readings from the multiprobe and the extraction procedure indicated that they were significantly correlated but temperature dependent.

Language

Identifier

Unique identifying numbers for this thesis in the Digital Library or other systems.

Collections

This thesis is part of the following collection of related materials.

UNT Theses and Dissertations

Theses and dissertations represent a wealth of scholarly and artistic content created by masters and doctoral students in the degree-seeking process. Some ETDs in this collection are restricted to use by the UNT community.

What responsibilities do I have when using this thesis?

When

Dates and time periods associated with this thesis.

Creation Date

  • May 2001

Added to The UNT Digital Library

  • Sept. 25, 2007, 9:26 p.m.

Description Last Updated

  • Dec. 12, 2013, 2:29 p.m.

Usage Statistics

When was this thesis last used?

Yesterday: 0
Past 30 days: 0
Total Uses: 1,455

Interact With This Thesis

Here are some suggestions for what to do next.

Start Reading

PDF Version Also Available for Download.

International Image Interoperability Framework

IIF Logo

We support the IIIF Presentation API

Lambert, Patricia. Evaluation of the Chlorophyll/Fluorescence Sensor of the YSI Multiprobe: Comparison to an Acetone Extraction Procedure, thesis, May 2001; Denton, Texas. (https://digital.library.unt.edu/ark:/67531/metadc2794/: accessed May 27, 2024), University of North Texas Libraries, UNT Digital Library, https://digital.library.unt.edu; .

Back to Top of Screen