The number of emitted photons that can be obtained from a fluorophore increases with the incident light intensity and the duration of illumination. However, saturation of the absorption transition and photodestruction place natural limits on the ultimate signal-to-noise ratio that can be obtained. Equations have been derived to describe the fluorescence-to-background-noise ratio in the presence of saturating light intensities and photodestruction. The fluorescence lifetime and the photodestruction quantum yield are the key parameters that determine the optimum light intensity and exposure time. To test this theory we have performed single molecule detection of phycoerythrin (PE). The laser power was selected …
continued below
Publisher Info:
Stanford Univ., CA (United States). Dept. of Cell Biology
Place of Publication:
California
Provided By
UNT Libraries Government Documents Department
Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.
Descriptive information to help identify this article.
Follow the links below to find similar items on the Digital Library.
Description
The number of emitted photons that can be obtained from a fluorophore increases with the incident light intensity and the duration of illumination. However, saturation of the absorption transition and photodestruction place natural limits on the ultimate signal-to-noise ratio that can be obtained. Equations have been derived to describe the fluorescence-to-background-noise ratio in the presence of saturating light intensities and photodestruction. The fluorescence lifetime and the photodestruction quantum yield are the key parameters that determine the optimum light intensity and exposure time. To test this theory we have performed single molecule detection of phycoerythrin (PE). The laser power was selected to give a mean time between absorptions approximately equal to the fluorescence decay rate. The transit time was selected to be nearly equal to the photodestruction time of {approximately}600 {mu}s. Under these conditions the photocount distribution function, the photocount autocorrelation function, and the concentration dependence clearly show that we are detecting bursts of fluorescence from individual fluorophores. A hard-wired version of this single-molecule detection system was used to measure the concentration of PE down to 10{sup {minus}15} M. This single-molecule counter is three orders-of-magnitude more sensitive than conventional fluorescence detection systems. The approach presented here should be useful in the optimization of fluorescence detected DNA sequencing gels. 17 refs., 4 figs.
This article is part of the following collection of related materials.
Office of Scientific & Technical Information Technical Reports
Reports, articles and other documents harvested from the Office of Scientific and Technical Information.
Office of Scientific and Technical Information (OSTI) is the Department of Energy (DOE) office that collects, preserves, and disseminates DOE-sponsored research and development (R&D) results that are the outcomes of R&D projects or other funded activities at DOE labs and facilities nationwide and grantees at universities and other institutions.
Mathies, R. A.; Peck, K. & Stryer, L.High-Sensitivity Single-Molecule Fluorescence Detection in Theory and Practice,