Mass-spectrometric determination of trace elements in aqueous media without preconcentration
Description:
Feasibility of using a low pressure glow discharge as an ion source for the mass spectrometric determination of trace elements in aqueous media was investigated. A cryogenically cooled hollow cathode ion source was developed to analyze aqueous samples without external preconcentration. Aqueous solutions containing seventy elements were analyzed and the detection limits, sensitivity factors, and linear regression correlation coefficients were determined. A standard test solution of trace element…